| Sensor |
NTC Thermistor
|
| Measurement RangeResolution |
-86°C to +35° recommended, may be used up to +55° 0.1°C
|
| Calibrated Accuracy: |
±1.0°C (over specified accuracy range)
|
| Specified Accuracy Range: |
-80°C to 0°C
|
| Readings: |
32,767 readings
|
| Recording Interval: |
5 seconds to 30 minutes, selectable in software
|
| Start Modes: |
Immediate start, delay start, manual start with magnetic switch
|
| Programmable Start Delay: |
Up to 30 days from PC launch or manual start
|
| Temperature Alarm: |
Configurable temperature alarm with programmable high and low alarm limits, high and low warn limits
|
|
Each set point may be individually enabled or disabled during acquisition
|
| Status Indicators: |
Green: Blinks to indicate safe conditions
|
|
Yellow: Blinks to indicate warn limits have been exceeded
|
|
Red: Blinks to indicate alarm criteria (limits/ delay)have been exceeded
|
| Password Protection: |
An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password.
|
| Calibration Date: |
Automatically recorded within device
|
| Power: |
3.6V lithium battery (LTC-7PN), factory replaceable (not user replaceable)
|
| Battery Life: |
2 years typical @ 25°C, 1 min. reading intervals 1 year typical @ –80°C, 1 min. reading intervals
|
| Time Accuracy: |
±2 minutes/month max. at 20 to 30°C (±100 ppm typical) ±5 minutes/month max at –80°C (±100 ppm typical)
|
| Data Format: |
Date and time stamped °C, °F, K, °R USB docking station to PC (IFC300) 38,400 baud
|
| Software: |
Windows 95/98/ME/NT/2000/XP based software
|
| Operating Environment: |
-86°C to +35°C
|
| Dimensions: |
3.3” x 2.1” x 0.4” (83 mm x 54 mm x 11 mm)
|
| Ingress Protection/IP Rating: |
IP64 (splash resistant, not submersible)
|
| Enclosure: |
ABS plastic, white, polycarbonate labeling
|
| Weight: |
1.4 oz (40 g)
|
| Compliance: |
CE
|