| Measurement Parameter |
Temperature
|
Humidity
|
Pressure
|
| Measurement Channel |
1 Channel
|
1 Channel
|
1 Channel
|
| Internal Sensor |
-10 to 60°C
|
|
750 to 1100 hPa
|
| Standard External Sensor |
0 to 50°C
|
10 to 95% RH
|
|
| Optional External Sensor |
-40 to 110°C
|
|
|
| Measurement Accuracy |
±0.3% (0 to 50°C)
±0.5% (-40 to -20 / 80 to 110°C)
|
±5% RH
(25°C, 50% RH)
|
±1.5% hPa
|
| Display Resolution |
0.1°C
|
1% RH
|
0.1 hPa
|
| Sensor Type |
Thermistor
|
Macromolecular
|
Atmospheric
Pressure Sensor
|
| Internal Sensor Time Constant |
12 Sec.
|
|
4 Sec. **1
|
| Recording Interval |
1, 2, 5, 10, 15, 20, 30 Seconds / 1, 2, 5, 10, 15, 20, 30, 60 Minutes- Total of 15 choices |
| Recording Capacity |
8,000 Readings × 2 or 3 Channels |
| Recording Mode |
Endless Loop Method (Overwrite from the oldest data when recording capacity is full)
One Time Method (Stop recording when recording capacity is full) |
| LCD Display |
Measurements, (Any Single Channel or alternating display of all Channels),
Recording Status, Battery Life Warning, Amount of Recorded Data, Unit of Measurement |
| Power |
AA alkaline battery (LR6) |
| Battery Life |
Approx. 10 months |
| Data Back-up |
Activated when battery power is low or when switch is off (Approx. 1 year) |
| Communication Interface |
USB Communication Cable (included) |
| Data Download Time |
Approximately 8 seconds (For full memory) |
| Dimensions |
H55mm × W78mm × D18mm |
| Weight |
About 62g (Including one AA battery) |
| Operating Environment |
-10 to 60°C, 90% RH (non-condensing) |
| Standard Sensor (Included) |
TR-3100 External Temperature & Humidity Sensor |
Accessories
included in package |
AA alkaline Battery (LR6), User's Manual,
USB Cable (US-15C ; length 1.5m), Software CD Set |
|
**1 Pressure time constant is 40 seconds for sample rates of one minute or longer.
|