£2,805.00

400°C temperature validation system. Including: ThermoVault Max thermal shield; HiTemp140 M12 temperature data logger and M12 Depyrogenation probe. For time limited operation at temperatures between –200°C to +400°C.

Sub-total
£2,805.00

Product Description

The DHS comprises of a ThermoVault Max thermal shield, HiTemp140 M12  temperature data logger and M12 Depyrogenation probe.

Durable stainless steel thermal shield from MadgeTech enables use of the supplied data logger at temperatures beyond the stated operational range for limited time periods. Please review the associated product brochure for details on the length of time the system can operate at temperatures between -200°C ~ -40°C and +140°C ~ +400°C or contact our sales team for advice.

Market leading solution for precise and validation of time vs temperature in dry heat sterilisation applications

Requires: IFC400 or IFC406 interface package.

 

 

 

Manufacturer

Madgetech

Parameter(s)

Temperature

Usage Type

Multiple

Connectivity

USB via Interface

Sensor internal or external

Internal

Temperature Sensor

36 inch Glass Braided RTD depyrogenation probe

Temperature Range

-200 °C to +400 °C

Temperature Resolution

0.0001 Ω, 0.01 °C (0.02 °F)

Temperature Calibrated Accuracy

±1.2 °C (-200 °C to +400 °C)

Memory

43,690 readings

Reading Rate

4 readings per second up to 1 reading every 24 hours

Baud Rate

125,000

Battery Life

2 years typical (1 minute reading rate at +25 °C / +77 °F) User Replaceable Battery

Material

316 Stainless Steel, PTFE, Silicone

Dimensions

1.75 in OD x 9.6 in L (44.45 mm OD x 243.8 mm L)

Operating Environment

-200 °C to 400 °C %RH to 95 %RH non-condensing

Required Interface Package

IFC400 or IFC406

Product Brochure
Dry Heat Sterilization System DS WPLS PDF 265 KB
HiTemp140 M12 Data Sheet WPLS PDF 998 KB
M12 Depyro Data Sheet WPLS PDF 254 KB
ThermoVault Max Data Sheet WPLS PDF 218 KB
Instruction Manual
ThermoVault Max User Guide WPLS PDF 2 MB
TMAX Wet Seal Kit User Guide WPLS PDF 1 MB

Product Description

The DHS comprises of a ThermoVault Max thermal shield, HiTemp140 M12  temperature data logger and M12 Depyrogenation probe.

Durable stainless steel thermal shield from MadgeTech enables use of the supplied data logger at temperatures beyond the stated operational range for limited time periods. Please review the associated product brochure for details on the length of time the system can operate at temperatures between -200°C ~ -40°C and +140°C ~ +400°C or contact our sales team for advice.

Market leading solution for precise and validation of time vs temperature in dry heat sterilisation applications

Requires: IFC400 or IFC406 interface package.

 

 

 

Manufacturer

Madgetech

Parameter(s)

Temperature

Usage Type

Multiple

Connectivity

USB via Interface

Sensor internal or external

Internal

Temperature Sensor

36 inch Glass Braided RTD depyrogenation probe

Temperature Range

-200 °C to +400 °C

Temperature Resolution

0.0001 Ω, 0.01 °C (0.02 °F)

Temperature Calibrated Accuracy

±1.2 °C (-200 °C to +400 °C)

Memory

43,690 readings

Reading Rate

4 readings per second up to 1 reading every 24 hours

Baud Rate

125,000

Battery Life

2 years typical (1 minute reading rate at +25 °C / +77 °F) User Replaceable Battery

Material

316 Stainless Steel, PTFE, Silicone

Dimensions

1.75 in OD x 9.6 in L (44.45 mm OD x 243.8 mm L)

Operating Environment

-200 °C to 400 °C %RH to 95 %RH non-condensing

Required Interface Package

IFC400 or IFC406

Product Brochure
Dry Heat Sterilization System DS WPLS PDF 265 KB
HiTemp140 M12 Data Sheet WPLS PDF 998 KB
M12 Depyro Data Sheet WPLS PDF 254 KB
ThermoVault Max Data Sheet WPLS PDF 218 KB
Instruction Manual
ThermoVault Max User Guide WPLS PDF 2 MB