HiTemp140-FRFastResponseExtremeTemperatureDataLogger

By MadgeTech Data Loggers
From: £460.00

Overview

The HiTemp 140 FR is an extreme data logger with a 2” temperature probe, designed for use in high temperature environments. Capable of fast responses, this temperature data logger has a 4Hz recording rate. Includes ISO17025 calibration at 30 and 140°C

Customise this datalogger

Part of a suite of products designed to work as a complete monitoring system

  • Custom 3 point calibration certificate
  • Data Interface - Required to download data
  • Spare Battery
* Calibration takes between 3 - 7 business days Add and customise

Measures

Accurate (up to 0.1 °C) and rapid (4Hz) recording of temperatures from -200 to + 260 °C via needle probe

Easy to use

Self contained operation with interface and user friendly software. 21 CFR Part 11 compliance and GxP documentation available

ISO17025 Calibration

Tested at +30 and +140 °C and supplied with certificate

Information

Product Description

Please Note : Required: IFC400 or IFC406 interface along with free issued MT4 software (see downloads section).

This fast-response temperature data logger is a unique solution for rapidly sampling and recording precise temperature data during validation routines in ovens, autoclaves and sterilisers. The durable construction allows for use in extreme temperature environments.

The HiTemp140-FR joins the ever-expanding range of MadgeTech data loggers for validation purposes, which encompass the demand for reliability and high performance at reasonable prices.

Includes ISO17025 calibration certificate at standard points 30 and 140°C

Guidance for using the HiTemp140-FR extreme data logger

Configure a single HiTemp140 FR using the IFC400 interface or multiple data loggers with the IFC406 multiplexer. This submersible data logger is made from stainless steel and has an IP68 construction, enabling operation in any environment between -40°C and +140°C (with no time limit). Use the TSK shield for limited-time use of the extreme data logger in -200°C to +250°C.

Once the validation exercise has been concluded, wait for the fast response data loggers to cool down or warm up and insert them back into the interface. Using the MT4 software and ‘workflows’ to automatically download and save data, you can generate customised reports and perform analytical functions.

HiTemp140 FR features include:

  • Recording of graphical temperature vs. time with threshold analysis
  • Generation of a table of temperature vs. time data
  • Statistical evaluation of:
    • Maximum
    • Minimum
    • Average
    • Change over time
    • F0
    • PU
  • Self-contained operation with optical interface is easy to use
  • User friendly temperature data logger software

Optional: Calibration Certificate

Calibration services ensure that fast response data loggers are calibrated to achieve specified accuracy. It’s recommended that a calibration certificate is obtained to prove temperature validation for 21 CFR Part 11 compliance.

Calibration services available:

  • Standard calibration: 30 and 140°C
  • Custom calibration services: 3 user defined points across measuring or operating range
  • ISO17025 calibration available on request

Specifications

Manufacturer

MadgeTech

Parameter(s)

Temperature

Usage Type

Multiple

Connectivity

USB via interface

Sensor internal or external

Internal

Channels

1

Sensor type

2" x 0.0625" diameter, 100 ohm Platinum RTD

Response time

Response Time Slow Moving Air 63% = 39 seconds
Response Time Slow Moving Air 90% = 163 seconds
Response Time Water 63% = 10 seconds
Response Time Water 90% = 12 seconds

Memory

65,536 readings

Sample rate

4 readings a second

Range

-200°C ~ + 260°C

Accuracy

+/- 0.1°C

Resolution

0.01°C

Material

316 Stainless Steel

Power

User replaceable battery

Battery Life

1 year typical (1 minute reading rate at 25 °C/77 °F)

Dimensions

Body - 1.89 in x 0.970 in x 0.970 in (48 mm x 24.6 mm x 24.6 mm)

Probe - 1.75 in x 0.0625 in dia. (0.188 in transitional dia.)
44.5 mm x 1.59 mm (4.78 mm Transitional dia.)

Environment

IP68

Operating Conditions

-40 ºC to +140 ºC (-40 ºF to +284 ºF)
0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA

Features

Rapid sampling and recording
Suitable for use in autoclaves
Compatible with thermal shield for use in extreme temperatures
21 CFR Part 11 Compliant
Precise temperature data logging
Batch configuration with multiplexer