LogTagGlycol
LogTag Glycol Buffer Adapter
Part of Wessex Power • Looking for Compliant Monitoring Solutions? Visit wessexpower.co.uk
Call 01929 459459 or email with any questions
The TRED30-16U is a min/max temperature data logger with audible alert and connector for an external temperature sensor, LCD display, 30-day summary, memory for 17’526 readings and user replaceable battery.
Enhance your monitoring with these options
Audible Alarm
Integrated audible alarm triggered if readings outside pre-set limits are breached
LCD Display
The display allows immediate viewing of current and min/max /alarm, daily entries over a 30 day period.
Eliminate calibration headaches with the plug and play smart probes
Digital calibrated probes. Hot-swappable, no need to remove loggers from current locations.
Requires LogTag temperature probe, USB cradle interface and LogTag software for measurement, configuration and analysis (see downloads section).
The TRED30-16U is ideally suited for use as a fridge or freezer data logger where temperatures of time temperature sensitive medical products are stored. It is also ideal in the validation of shipping containers used to transport medicinal products in the supply chain.
The easy to read LCD display delivers key information at a glance and enables users to navigate through a 30 day summary of min/max results and alarm events before interfacing with a computer.
Display functions: Current temperature | Alarm Status - In range (Tick) out of range (cross) | Recording status| Battery condition
If a measurement is recorded outside of user definable temperature alarm limits an audible alert will be generated to attract the attention of responsible personnel.
The next step will be to review details of the alarm on the local LCD display before interfacing the data logger with a computer to inspect the collected time vs temperature data.
Configuration is performed using LogTag data logger software. Pair and address the LogTag TRED30-16U with an appliance or provide a unique name for use in mapping the thermal properties of an appliance or location within a storage area. Set a recording interval, low and high alarm limits and the method of triggering the recording (Button start or predefined date and time) and deploy the data logger using the handy anchoring point (Hole in the corner) or using the optional mounting bracket. Review the data using the LCD display and collect the data logger for download in the event of temperature excursion or for weekly or monthly audit. Upon insertion of the data logger into the USB cradle, downloaded data is stored automatically and presented in a clear and concise report which can be printed or distributed via email. Additional analysis can be performed within the software using the graphical, tabular and statistical features.
Supplied with user replaceable battery.
LogTag
Temperature
Multiple Use
USB via interface
External
1
Thermistor
Capacity 17,526 temperature readings
60 days @ 5 minute
30 days of Min/Max/Duration and alarm statistics
Interval Configurable from 30 seconds to 18
-40°C ~ +99°C
Better than ±0.5 °C (±0.9 °F) for -20 °C to +40 °C (-4° F to +104 °F), typically ±0.3 °C (±0.6 °F)
Better than ±0.8 °C (±1.4 °F) for -40 °C to -20 °C (-40 °F to -4 °F), typically ±0.5 °C (±0.9 °F)
Better than ±0.8 °C (±1.4 °F) for +40 °C to +70 °C (+104 °F to +158 °F), typically ±0.7 °C (±1.3 °F)
Better than ±1.2 °C (±2.2 °F) for +70 °C to +99 °C (+158 °F to +210 °F), typically ±1 °C (±1.8 °
Contact our calibration department for advice on offset adjustment.
0.1 °C (0.2 °F) for measurements -40 °C to +50 °C (-40 °F to +122 °F)
0.2 °C (0.4 °F) for measurements +50 °C to +70 °C (+122 °F to +158 °F)
0.3 °C (0.5 °F) for measurements +70 °C to +80 °C (+158 °F to +176 °F)
0.4 °C to 0.6 °C (+0.7 °F to +1 °F) for measurements +80 °C to +99 °C (+176 °F to +210 °
Polycarbonate
3V CR2032 (LiMnO2) coin cell battery, user replaceable
12 months
Assumes 6 minute recording interval, daily review of Min/Max on LCD and daily download
NOTE: Continual activation of the audible alarm will reduce the working life of the battery.
93mm (H) x 54.5mm (W) x 8.6mm (T)
IP61 (when vertically mounted or hung)
-30°C ~ +60°C
Audible
LCD
Detailed in PDF Report
Audible alert, LCD and button 30 day calendar Min/Max, alarm summary navigation feature
Windows PC
RTCA/DO‐160G section 21 as category H
LogTag Glycol Buffer Adapter
65mm medium tipped probe, 3.2mm diameter with 1.5M of flat...
© Loggershop Copyright is property of Wessex Power Technology Copyright 2026
Privacy Policy | Contact Us | Calibration Enquiry | Terms & Conditions
ISO9001 Certified for customer service 


